Hardware Specifications:
- DIMENSIONS - 4.3 x 12.5 x 11.6 inches (10.9 x 31.8 x 27.5 cm)
- WEIGHT - 12.8 lbs. (5.8 kg)
- COMMUNICATIONS
- PROTOCOL - ASCII Standard
- DATA TRASFER RATE - 9600 Baud
- POWER - 110/220 VAC or rechargeable battery (approximately
4 hours per charge)
- CHANNEL SIZES - Eight user selectable
- SIZE RANGE - 2µm-400µm standard,
1.3 and 1.0 micron options available
- SENSOR, STANDARD - Resolution 5% or less at 10
micron and 9% or less at 2 micron per ASTM-F-658 on Std # LD400
with a concentration
limit better than 18,000 counts/ml
- SENSOR, LARGE BORE- Optional for upstream settled and raw water
monitoring for concentrations up to 35,000 counts/ml, orifice
size 1 or 2 mm in diameter
- SIGNAL TO NOISE - Greater than 3 at lowest micron rating
- COUNTING EFFICIENCY - 50%@ 1.0 micron, >98%@ 1.5 micron, > 99.9%@
2.0 micron
- MEMORY - 300 Sample runs
- LIGHT SOURCE - Laser Diode
- FLOW RATE - 25 mL/min, 60 mL/min standard, 100 mL/min and other
flow rates available
- DATA TRANSFER - User selectable baud rate - 1200, 2400, 4800
or 9600
- DATA ENTRY - Integrated bar code entry
- ALARMS - User selectable alarm limits/audible alarms
|
Software and Computer
Requirements:
- OPERATING SYSTEM - Windows 3.1, Windows 95/98, Windows NT,
Windows XP Pro
- COMMUNICATION - Requires free RS-232 serial port
- COMPUTER - Pentium 133MHz with 16MB RAM or better
|